Vacuum hot pressing furnace use Common Faults and exclusion method

High vacuum pumping on but does not leak (mean value of the vacuum table for a long time does not move)
Possible causes:
1. High vacuum gauge tube aging
 2. Diffusion pump oil aging
Solution: Replace the vacuum gauge; replacement diffusion pump oil.
But not on the low vacuum pumping does not leak (mean value of the vacuum table for a long time does not move)
Possible causes:
1 low vacuum current regulatory work incorrectly.
2. Mechanical pump oil aging
Solution: Adjust vacuum gauge tube current; replacement of mechanical pump oil.
Thermocouple automatic access failure, warning lights alarm
Possible causes:
1. Inner and outer limit failure.
2. Out of the motor does not turn.
Solution: Adjust or replace the external stopper; tests whether the motor has been broken, such as the replacement of the same type of motor bad. (Note) within the furnace limit should be observed before heating indicator light should be lit.
After a period of time the temperature does not rise up
Possible causes: heater resistance value becomes larger.
Solution: According to instructions to transfer large instrument output power; serious aging can only be replaced with a new heater
Incase control panel lights are on and the police overcurrent
Possible causes: heater shorted
Solution: Open the lid heater observe whether there is a short circuit, troubleshooting.
Over-temperature alarm and alarm lights on
Possible causes:
1. Over-temperature furnace
2. Instrument is set incorrectly
Solution: Press the manual reset over-temperature alarm parameters
Without water is lit and the alarm
Possible causes:
1. Valve did not open
2. Water pressure is too low
Solution: Open the valve; increase water pressure.